Spinal Surgery News return to home page
    For Today's Orthopaedic Professional

ECiMa™ HXLPE for low wear and oxidation resistance of hip bearings
Author: Alison Traynor MEng, David Simpson BEng PhD and Simon Collins BSc PhD CEng FIMechE. Corin, UK

Pushing Orthopaedics: Independent research from the Implant Development Centre Orthopaedics


The association between hip fracture and hip osteoarthritis: A case-control study
Authors: Jonas Franklin1,2, Martin Englund2,3, Torvaldur Ingvarsson1,4 and Stefan Lohmander2

1University Hospital, Akureyri, Iceland
2Department of Orthopedics, Clinical Sciences Lund, Lund University, Sweden
3Clinical Epidemiology Research & Training Unit, Boston University School of Medicine, Boston, MA, USA
4Department of Health Sciences, University of Akureyri, Iceland and Faculty of Medicine, University of Iceland, Reykjavík, Iceland

BMC Musculoskeletal Disorders 2010, 11:274doi:10.1186/1471-2474-11-274
© 2010 Franklin et al; licensee BioMed Central Ltd.




ECiMa™ HXLPE for low wear and oxidation resistance of hip bearings
Author: Alison Traynor MEng, David Simpson BEng PhD and Simon Collins BSc PhD CEng FIMechE. Corin, UK

Introduction
Ultra high molecular weight polyethylene (UHMWPE) remains a widely used bearing in total joint replacement. Wear induced osteolysis remains a concern with such bearings and first generation highly crosslinked polyethylene (HXLPE) materials were introduced to address this issue [1]. However, there are concerns regarding bulk material property degradation of HXLPE [2] and oxidation after implantation which may be as a result of lipid absorption or cyclic loading [3].

ECiMa is a cold-irradiated, mechanically annealed a-tocopherol (vitaminE) blended UHMWPE and is a next generation HXLPE developed to maintain mechanical properties, minimise wear and to improve the oxidation resistance in the long-term.

The aim of this study was to compare the in-vitro wear rate and mechanical properties of three different UHMWPE acetabular liners; conventional UHMWPE, HXLPE and ECiMa.

Methods
Fifteen acetabular liners (Corin, UK) underwent a hip simulation (servo-hydraulic, type C6/2-07). Three conventional UHMWPE liners(GUR 1050, Ø32 mm, gamma sterilised in nitrogen to 30kGy) were tested to 3 million cycles (mc). Six HXLPE liners (GUR 1020, Ø40 mm, cross-linked to an irradiation dose of 75kGy and EtO sterilised) and six ECiMa liners (GUR 1020 blended with 0.1wt% vitamin E, Ø40 mm, cross-linked to an irradiation dose of 120kGy, mechanically deformed and annealed, and EtO sterilised) were tested to 5mc with all liners articulating against CoCrMo alloy femoral heads.

The wear simulator used a double-peak force curve [4], with a maximum force of 3.0 kN. The test lubricant used was calf serum with a protein content of 30 g/l supplemented with 1% (v/v) patricin as an antibacterial agent. The simulators operated at a frequency of 1 Hz and were stopped after the first 0.5 mc and every 1.0 mc thereafter; volumetric wear rate was determined gravimetrically. Moisture uptake was monitored using loaded soak controls; one for UHMWPE and two for HXLPE and ECiMa.

Three of the ECiMa liners were cut in half post wear testing. One half of each was subject to accelerated ageing (ASTM F2003-02, 5atm of pure oxygen at 70ºC for 14 days), while the other half was tested as received. Slices, 200 µm thick, were taken from each liner half. Hexane extraction was performed on both the aged and unaged samples prior to oxidation analysis to remove any lipid contaminants from the wear testing. Oxidation analysis was performed using a Fourier Transform Infra-red technique (ASTM F2102-01) throughout the thickness of each liner half. Average oxidation indices for each sample were determined.

To provide comparative data, samples of untested ECiMa material (n=3) were aged in accordance with ASTM F2003-02.

Dog bone specimens, 3.2 mm thick (ASTM D638 type V) were machined from ECiMa material. Half of the samples were aged in accordance with ASTM F2003-02. Uniaxial tension testing was carried out as per ASTM D638 to characterise the mechanical properties of the aged and unaged material. Specimens were machined in three different directions; parallel to the direction of consolidation, parallel tothe direction of deformation and orthogonal to both of these directions.

Ultimate tensile strength (UTS), yield strength and percent elongation values were measured. These values were compared to mechanical data available for the other material types. A further set of dog bone specimens (1.8 mm thick, ASTM D638 type IV) were machined from HXLPE. Half of the samples were aged in accordance with ASTM F2003-02. All samples were tested to characterise the mechanical properties. These values have also been compared to mechanical data available for conventional UHMWPE.

Figure 1: Comparison of wear rate (mm3/mc) for the 3 different materials (5mc data for HXLPE and ECiMa)
Results and discussion
There was a 95% and a 83% reduction in the wear rate for the ECiMa liners compared to the conventional UHMWPE and HXLPE liners respectively (Figure 1).

All of the oxidation values for the wear tested ECiMa liners, before and after ageing, and the aged, untested ECiMa samples were negative and below the level of detection. This indicates a high level of throughthickness oxidation resistance even after aggressive ageing and cyclic loading.

Table 1: Comparison of Mechanical Properties Note: ECiMa data presented is the average of the three directions tested
The data indicates an increase in UTS, yield strength and percent elongation of 12%, 10% and 8% respectively, for unaged ECiMa compared to HXLPE. Importantly, the mechanical properties were not substantially degraded when compared to conventional UHMWPE and superior to the HXLPE samples.

Conclusion
The reduced wear rate during in-vitro hip simulation of ECiMa compared to conventional UHMWPE, coupled with improved mechanical properties and long-term oxidation resistance in comparison to HXLPE, indicates ECiMa as a promising next generation UHMWPE bearing material.

References
  1. Jacobs et al. JBJS, 2007, 86 (12): 2779-2786
  2. Oral et al. Biomaterials, 2006, 27: 917-925
  3. Rowell et al. ORS 2010: Paper #358
  4. Paul J.P. IMechE Proc. H, 1966, 181 (3J): 8-15
  5. Dumbleton et al. CORR, 2006, 453:265-271


top  | next
Pushing Orthopaedics: Independent research from the Implant Development Centre Orthopaedics


Research at the heart of the Implant Development Centre
At the heart of the Implant Development Centre (IDC), one of the technology hubs for Smith & Nephew, is the desire to advance the understanding of how implants perform in vivo in hips.

It all started back during the design conception of the BIRMINGHAM HIP™ Resurfacing system, which was developed based on forensic analysis of retrieved historical metal-on-metal devices.

Following this tradition, the Implant Development Centre has become a worldwide recognised research centre with its state-of-the-art facilities equipped with cutting edge scientific equipment. The IDC is recognised as one of the most advanced hip arthroplasty research facilities in the world.

The understanding gathered through the research is then applied to solutions to better improve the quality of life of the patients.

Independent accredited research
This research would be pointless unless it is structured to be independent, unbiased and scientific.

A number of accreditations certify the independency of our research.

The IDC is the only UKAS accredited hip wear and friction simulator testing centre in the world.

The IDC is also ILAC accredited – ILAC is an international cooperation of laboratory and inspection accreditation bodies formed more than 30 years ago.

These accreditations ensures that the hip wear and friction simulator studies carried out at the IDC have been assessed against internationally recognised standards to demonstrate the centre’s technical competence, impartiality and performance capability.

The Research
A modern hip simulator test cell with reservoir of lubricant (Bovine Serum) containing implants being tested.
Our current aims at the IDC are:
  • provide a retrieval analysis service for all hip arthroplasty surgeons
  • provide scientific tests that replicate clinical environments
  • publish regularly our scientific findings at national, international and global meetings
  • undertake retrospective scientific studies to further understand clinical outcomes
  • study relevant hip biomechanics, kinetics and kinematics
  • Synthesis and investigation of new biomaterials, potentially useful in joint arthroplasty
  • Facilitating partnership with surgeons and other parties to further these areas of interest
  • Enhancement and validation of current manufacturing techniques.
Achievements
Example of a patient walking on a treadmill to simulate running speed of hip simulator.
The IDC has over 40 publications in either peer reviewed journals, conference presentations and posters, as well as having contributed to a number of orthopaedic text books.

One of the most powerful investigations coming out of the Implant Development Centre is the development of a physiological relevant hip simulator protocol.

The use of hip simulators to validate the performance of prosthesis has frequently highlighted disparities between the results from the laboratory and the performance in-vivo.

In some circumstances the hip simulators may artificially produce an exaggerated lubrication regime and potentially underestimate the amount of wear and that would explain these disparities the between in-vivo and in-vitro settings.

The physiologically relevant test protocol developed at the IDC reproduces daily activities, stop/start motion, and various kinetics and kinematics that may break down the fluid film to provide a realistic picture of how an implant might behave in-vivo.

The validity of the physiological relevant test protocol was validated for first time when it demonstrated in- vitro the difference in wear between “heat treated” vs “as cast” bearings. These results more accurately correlate to clinical metal ion data comparing “as cast” and “heat treated” devices.

Other IDC studies:
  • The tribology performance and wear of heat treated vs as-cast CoCr metallurgy
  • The wear of low clearance MoM bearings in hip simulator contrasted with clinical results
  • The tribiological performance using a worn BHR acetabular cup with a new femoral head in cases of femoral neck fracture
  • The effect of acetabular cup orientation on wear in hip simulators
  • Friction experienced in various bearings using different physiologically relevant lubricants
  • The clinical effect of acetabular cup deformation and deflection on cup insertion
  • The effect of microstructure on wear characteristics
The validity of the work carried out at the Implant Development Centre is endorsed by substantial contributions to a number of major orthopaedic books, including “Modern Hip Resurfacing “ by Derek McMinn with several chapters on hip joint tribology, corrosion and its contribution to metal release and retrieval analysis.

Last year the IDC was approached by EFORT to contribute in their forthcoming “Tribology in Total Hip Arthroplasty” book with a chapter devoted to the tribology of Metal on Metal bearings (chapter 4).

Next year will see the publication of “Mastering Orthopaedic Techniques in Total Hip Arthroplasty” in which the editors also approached the IDC to contribute to a full chapter devoted to tribology and bearing materials (chapter 2).

The Retrieval analysis project
From the design conception of the BIRMINGHAM HIP™ Resurfacing system, forensic analysis of retrieved historical metal-on-metal devices was an integral piece of the puzzle when gaining an understanding of the recipe to success.

From the intimate comprehension of this process and continued pursuit for insight, the retrieval analysis project was initiated as a scientific resource in July 2000 with the intention of studying retrieved hip arthroplasty devices to gain a further insight into their function and performance within the body.

It now has one of the largest explant collections in the world that are analysed with state of the art equipment (SEM with EDX for materials analysis, Optical microscopes, Taylor Hobson roundness profiler, Redlux 3D profiler, Surface roughness profiler etc)

The IDC offers to the orthopaedic community explant testing and implant analysis service with several, reliable, replicable, accurate and standardised examination methods to investigate the explants including:
  • Metrology – including roundness and linear wear, surface analysis, dimensions, geometries and clearances
  • Material composition and microstructure: using cutting edge equipment such as scanning electron microscope (SEM) and Energy dispersive x-ray (EDX)
  • Radiographic analysis – 3d component alignment/positioning of the device
  • Histology
It is evident that at the heart of the Implant Development Centre is the desire to push the boundaries of orthopaedic research. It is also evident that the dedication and devotion of all the individuals involved in it and the quality of their work that makes possible the development of future advanced bearing technologies and developing knowledge, by forming partnerships within the medical and scientific community to achieve common goals.

top  | next
The association between hip fracture and hip osteoarthritis: A case-control study
Authors: Jonas Franklin1,2, Martin Englund2,3, Torvaldur Ingvarsson1,4 and Stefan Lohmander2

1University Hospital, Akureyri, Iceland
2Department of Orthopedics, Clinical Sciences Lund, Lund University, Sweden
3Clinical Epidemiology Research & Training Unit, Boston University School of Medicine, Boston, MA, USA
4Department of Health Sciences, University of Akureyri, Iceland and Faculty of Medicine, University of Iceland, Reykjavík, Iceland

BMC Musculoskeletal Disorders 2010, 11:274doi:10.1186/1471-2474-11-274
© 2010 Franklin et al; licensee BioMed Central Ltd.

Abstract
Background
There have been reports both supporting and refuting an inverse relationship between hip fracture and hip osteoarthritis (OA). We explore this relationship using a case-control study design.

Methods
Exclusion criteria were previous hip fracture (same side or contralateral side), age younger than 60 years, foreign nationality, pathological fracture, rheumatoid arthritis and cases were radiographic examinations were not found in the archives. We studied all subjects with hip fracture that remained after the exclusion process that were treated at Akureyri University Hospital, Iceland 1990-2008, n = 562 (74% women). Hip fracture cases were compared with a cohort of subjects with colon radiographs, n = 803 (54% women) to determine expected population prevalence of hip OA. Presence of radiographic hip OA was defined as a minimum joint space of 2.5 mm or less on an anteroposterior radiograph, or Kellgren and Lawrence grade 2 or higher. Possible causes of secondary osteoporosis were identified by review of medical records.

Results
The age-adjusted odds ratio (OR) for subjects with hip fracture having radiographic hip OA was 0.30 (95% confidence interval [95% CI] 0.12-0.74) for men and 0.33 (95% CI 0.19-0.58) for women, compared to controls. The probability for subjects with hip fracture and hip OA having a secondary cause of osteoporosis was three times higher than for subjects with hip fracture without hip OA.

Conclusion
The results of our study support an inverse relationship between hip fractures and hip OA.
Background
It is a common clinical observation that patients with hip fracture very rarely have hip osteoarthritis (OA)1-3. This has been examined in several studies, some claiming that patients with hip fracture have less hip OA than expected4-6, others that there is no difference between hip fracture patients and the general population7,8. Some studies have claimed that hip OA is only protective against intracapsular fractures9,10, while one study found that patients with hip OA have an increased risk for fracture11. An increased bone density in the femoral neck and a reduced density in the trochanter region in hips with OA, compared to hips without OA, was suggested to increase the risk for extracapsular fractures12. Low bone density increases the risk for hip fractures13, and patients with hip OA have higher bone density in the femoral neck14. An inverse relationship between hip OA and hip fracture has been suggested, possibly associated with genetic variation14,15.

The studies done so far have varied greatly in their methodology and definition of hip OA, some using self-report of hip OA4,5,8, and others a radiographic definition6,9-11,16. The latter studies have in addition used different scoring systems for the definition of hip OA. Some studies have been longitudinal and others cross-sectional. The cross-sectional studies have varied in the characteristics of the control groups. All this makes interpretation of the published evidence difficult.

The purpose of the present study was to test the hypothesis that subjects with hip fracture are less likely to have radiographic hip OA than control subjects without hip fracture.

Methods
The study was approved by the Ethics Committee of Akureyri University Hospital, where the study was based.

Figure 1: Flow chart of recruitment of hip fracture cases. n = the total number of subjects in each group, the number within parenthesis is the number of women in each group.
Hip fracture cases
In Iceland, as in other Scandinavian countries, all persons have a unique personal identification number. This, combined with a highly computerised national health system, makes it possible to identify all patients operated on for a given diagnosis or by a certain procedure in Iceland17. We identified 806 consecutive cases of hip fracture that were admitted to Akureyri University Hospital during 1990-2008. This constitutes 24% of all hip fractures in Iceland during that time. Patients were from both rural and urban areas. As younger hip fracture patients are more likely to have sustained a high energy trauma and sustain fracture in the absence of osteoporosis, all patients younger than 60 years and all patients with a previous hip fracture were excluded. After exclusions (Figure 1), we had 636 eligible cases with a first time hip fracture, and 562 of these had pre-operative hip radiographs available.

Controls with colon radiographs
Our control cohort was based on patients who had colon radiographs taken due to referral in routine health care. Colon radiographs taken at three different radiographic departments in Iceland during the years 1980-97 were examined. The patients were referred to these radiographic departments from four different hospitals (community and academic), as well as from the primary health care system. Patients were from both rural and urban areas. We excluded all subjects who were younger than 60 years of age, leaving 803 controls, of which 438 were women (54.5%). Iceland has a homogenous population of just over 300,000 inhabitants. The hip fracture patients were all treated at one hospital, but the controls came from three different hospitals. 38% of the controls were from the same hospital as the cases.

Radiographic techniques
The double contrast (barium enema) colon radiographs included at least two anteroposterior (AP) and several oblique exposures. In this study the hip joints were assessed from a supine AP control radiograph, which was taken with the same tube to film distance of 100 cm that is used in a standard AP view of the pelvis. The x-ray beam was centered on the umbilicus. Both hips were graded and we then randomly chose the right or the left hip for our control cohort.

In our case cohort preoperative radiographs were available for 562 (88%) cases. Pelvic radiographs showing the contralateral hip were available for 457 of patients (81.3%). The hip joints were assessed from a standard supine AP pelvic radiograph, taken with a tube to film distance of 100 cm with the x-ray beam centered on the symphysis pubis. We measured hip joint space with an electronic caliper18 and assessed osteophytes, sclerosis, cysts and any signs of secondary OA19. The same pre-operative radiographs were used for diagnosing the fracture and for the assessment of hip OA.

Hip fractures were classified as intra- or extracapsular20. We measured minimal joint space (MJS) and graded the radiographic features of OA according to Kellgren and Lawrence19, and individual features of OA, osteophytes, sclerosis and cysts in both hips.

All radiographs of cases were examined by a single observer (JF). All controls had been previously read by a different author (TI)21. To calculate inter- and intraobserver reliability22, we selected 50 radiographs of cases and 50 radiographs of controls, that included the full range of radiographic features that were read again by both these authors. We used Cohen’s kappa to calculate the intra- and interobserver reliability for detecting hip OA.

For MJS both the intra- and interobserver reliability was high, but Kellgren and Lawrence grading was not as reliable (Table 1).

Table 1. Intra- and interobserver reliability measured by Cohen’s ?
Validity of pre-operative radiographs
Pre-operative radiographs of cases were used since postoperative radiographs are now often taken in the operating theatre which yields radiographs that cannot be used to evaluate MJS. Presence of a hip fracture (displaced or not displaced) may affect the measurement of the joint space in the hip, even if subtle changes in position do not affect MJS in the hip23. To evaluate this we selected 50 patients where we had both pre- and postoperative standard pelvic radiographs, with an anatomic reduction of the fracture, and compared MJS measurement pre- and post-operatively. All the post-operative radiographs were taken within 10 days from the fracture. The 50 patients that were used for comparison of pre- and post-operative x-rays were not randomly chosen, since we were limited to patients that had a valid postoperative radiograph. We therefore tested if our sample adequately represented our fracture cases. Independent samples t-test for MJS was not significant (p = 0.96) and neither was Kolmogorov-Smirnov Z for OA/Not OA status (p = 0.36), which shows that this sample of 50 cases adequately represented our fracture cohort. There were 3 patients classified as having OA on a preoperative x-ray that were not classified as having OA on the postoperative x-ray. One patient was classified as having OA on the postoperative x-ray, that did not have OA on the preoperative x-ray. We calculated the p-value for pre- and postoperative readings of MJS with paired t-test (p = 0.18) and used the McNemars test to compare OA/Not OA status on pre- and postoperative x-rays (p = 0.63). Neither was significant, thereby validating the choice of using pre-operative x-ray examinations.

Unless otherwise stated, hip OA was defined as MJS of 2.5 mm or less21,22. In the hip fracture cohort there were 6 women and one man that had a total hip replacement due to OA in the contralateral hip. These were classified as having OA in that hip.

Secondary osteoporosis
Medications or diseases that cause secondary osteoporosis and thus increase the risk for hip fracture might be more common in the case or control group. To explore the association with risk factors for secondary osteoporosis in patients with hip fracture we did a prevalence study of possible risk factors for secondary osteoporosis within the hip fracture cohort using a matched design. This method was chosen as we did not have access to data on possible causes of secondary osteoporosis for the control cohort. We defined patients with hip fracture and MJS 2.5 mm or less as hip OA cases and to each of these we assigned two reference subjects, who had hip fracture, but not having hip OA (hence both hip OA cases and reference subjects came from the hip fracture cohort). The reference subjects were matched according to gender, fracture type and age ± 1 year. If more than 2 possible reference subjects were found, we randomly chose which to use. We reviewed the medical records of hip OA cases and their matched reference subjects from the fracture cohort and registered possible causes of secondary osteoporosis that we identified24,25.

Statistical methods
Age differences between groups were tested with the independent samples t-test. We age standardised the observed prevalence of hip OA in both hip fracture cases and controls against the Icelandic population from the year 2000 (using 5 year age strata and direct external standardisation)26. Intra- and interobserver reliability was calculated using Cohen’s Kappa. Paired t-test and McNemar test were used to compare readings of pre- and postoperative x-rays and we used independent samples t-test and Kolmogorov-Smirnov Z to see that this not randomly selected sample had approximately the same mean MJS and distribution of values as the entire group of cases. Chi-square was used in calculations for difference in crude rate and binary logistic regression for age-adjusted calculation of odds ratios for having OA amongst cases vs. controls. We considered a p-value of less than or equal to 0.05 to be significant, and all tests were 2-tailed. Calculations were done using SPSS v. 16.0. (SPSS Incorporated 2007).

Results
Case and control demographics and characteristics
The male to female ratio was about 1:3 in the fracture cohort, while it was close to one in the control cohort (Table 2). The number of basocervical and subtrochanteric fractures was small, so these were grouped together with pertrochanteric fractures as extracapsular fractures (Figure 1). There was no difference in mean age between the two fracture types, neither for men (p = 0.85), nor women (p = 0.48). The mean age of controls was significantly lower than that of subjects with fracture (p < 0.0001).

Table 2. Demographics of cases with hip fracture and controls
Prevalence of hip osteoarthritis
There was a significant difference in the crude prevalence of hip OA between fracture cases and controls for men (p = 0.02), but not for women (p = 0.3). Following a direct external standardisation of the crude prevalence, the differences between our cases and controls increased, especially for women where the age difference between cases and controls was greater (Table 3).

Table 3. Prevalence of radiographic hip OA in hip fracture cases and controls
Age-adjusted association between hip fracture and hip osteoarthritis
There was no significant difference in mean age when comparing men with or without hip OA (Table 4). However, women with hip OA were significantly older than women without hip OA.

Table 4. Mean age according to hip OA status
The age difference between cases and controls emphasised the need to take that into account when assessing the association between hip fracture and hip OA. We therefore did an age-adjusted binary logistic regression with hip OA as the dependent variable. All groups of hip fracture patients had a significantly reduced odds ratio of hip OA (comparing cases with controls) regardless of gender and classification system of hip OA. The odds ratio for hip OA was significantly reduced in both the fractured hip and the contralateral hip (Figure 2 and 3). For the fractured hip in men the OR was 0.30 (95%CI 0.12-0.74) when defining OA by MJS and 0.31 (95% CI 0.12-0.76) when defining OA by Kellgren and Lawrence. Corresponding results for the fractured hip in women were 0.33 (95%CI 0.19-0.58) and 0.38 (95% CI 0.21-0.69) when defining OA by Kellgren and Lawrence. For the contralateral hip in men the OR was 0.38 (95%CI 0.16-0.90) when defining OA by MJS and 0.38 (95% CI 0.16-0.91) when defining OA by Kellgren and Lawrence. Corresponding results for the contralateral hip in women were 0.55 (95%CI 0.32-0.96) and 0.52 (95% CI 0.29-0.94) when defining OA by Kellgren and Lawrence.

Figure 2. Age adjusted odds ratio for having hip OA in the fractured hip compared to controls without hip fracture. Figure 3. Age adjusted odds ratio for having hip OA in the contralateral hip of hip fracture cases compared to controls without hip fracture.
Error bars show 95% CI for cases compared to controls. MJS: Using hip OA defined by minimal joint space. K & L: Using hip OA defined by Kellgren and Lawrence grade. Error bars show 95% CI for cases compared to controls. MJS: Using hip OA defined by minimal joint space. K & L: Using hip OA defined by Kellgren and Lawrence grade.

Alternate definition of radiographic OA
In this study we have primarily used MJS 2.5 mm or less as definition of OA21,22, but used the Kellgren and Lawrence classification as well to enable comparison to previously published studies. We also tested our data using MJS 2.0 mm or less as definition of OA. The ORs were lower (meaning a greater difference between cases and controls), although the difference from MJS 2.5 mm or less definition was not statistically significant. Using MJS 2.0 mm or less as definition the crude prevalence of OA was 9.3% for male controls, 3.1% for male cases, 8.9% for female controls and 3.7% for female cases. For hip OA in men with hip fracture, we obtained the OR 0.18 (95%CI 0.06-0.57) and in women with hip fracture the OR 0.12 (0.05-0.28).

Intra- vs. extracapsular fractures
When defining hip OA by MJS we found no statistical difference in the age adjusted estimate of association for hip OA between the two different fracture groups, neither for men (OR = 0.96, 95% CI 0.21-4.5), or for women (OR = 1.5, 95% CI 0.75-2.8). When using Kellgren and Lawrence definition of OA, we again found no difference between the fracture types for men (OR = 0.96, 95% CI 0.21-4.5), but a significant difference for women, with extracapsular more frequently having hip OA than intracapsular fractures (OR = 2.5, 95% CI 1.2-5.2). We then repeated the calculations, using presence of OA in the contralateral hip. Again, there was no difference for men (OR = 2.3, 95% CI 0.52-10.4), regardless of classification system for OA. For women there was no difference for OA classified by MJS (OR = 1.8, 95% CI 0.92-3.5) or by Kellgren and Lawrence (OR = 1.5, 95% CI 0.73-3.2).

Secondary osteoporosis
There were 40 women and seven men who had hip fracture and hip OA in the same fractured hip according to definition by MJS. We reviewed the medical history for all, except one woman whose patient records were not found. We thus had 46 cases with hip fracture and hip OA matched by age, gender and fracture type to 92 reference subjects with hip fracture without hip OA. Patients were defined as having possible secondary osteoporosis if they had at least one of the conditions listed in Table 5. Three of 7 men and 15 of 39 women had at least one potential cause of secondary osteoporosis (Table 5). The probability of having a possible cause of secondary osteoporosis was 3 times higher for those with hip fracture and hip OA than for those with hip fracture, but without hip OA.

Table 5. Number (%) of hip OA cases and reference subjects without hip OA in the hip fracture cohort with risk factors for secondary osteoporosis
Discussion
The primary aim of this study was to evaluate the prevalence of hip OA in patients with hip fracture, compared to controls having had colon radiography. After adjusting for age, the odds ratio for hip OA in patients with hip fracture was found to be one-third of that in the comparison group. The difference in odds ratio between cases and controls was slightly less for the contralateral hip in women, but nevertheless significant. This suggests that the inverse relationship between hip OA and hip fracture is mainly systemic, i.e. affecting the whole patient, but contribution by a local effect in the arthritic hip cannot be excluded. The nature of the systemic effect is unknown, but a genetic factor may be involved.

For men there was no difference in age-adjusted prevalence of hip OA between those with intra- and extracapsular fractures, regardless of classification system and side. For women, there was a significantly greater prevalence of hip OA in extracapsular fractures only when defined by Kellgren and Lawrence in the fractured hip. When examining hip OA prevalence defined by MJS or when examining the contralateral hip, the difference between fracture types was not significant. We therefore conclude that there is no significant difference in hip OA prevalence when comparing intracapsular and extracapsular fractures. Our finding of a statistically significant difference for women in the fractured hip, when using Kellgren and Lawrence classification, might be a chance finding or perhaps the Kellgren and Lawrence classification system is less applicable to women, as was suggested27.

The probability for subjects with both hip fracture and hip OA having a possible secondary cause of osteoporosis was 3 times higher than for subjects with hip fracture but without hip OA. This finding supports the possibility that many of the patients with both hip fracture and hip OA had their hip fracture due to secondary osteoporosis. This could mean that if it had been possible to adjust for risk factors for secondary osteoporosis we might have found that the inverse relationship between hip fracture and hip OA is even stronger than reported in this study. These findings need to be interpreted with some caution as no gold standard definition of secondary osteoporosis exists and we retrieved information on secondary osteoporosis noted in medical records only, as we did not have bone density measures available. Further studies are needed to explore this aspect.

Possible confounders not accounted for in the present study are body mass index (BMI) and occupation, which have been shown to be independent risk factors for hip OA28, and use of medications, such as hormone replacement therapy and bisphosphonates, that affect osteoporosis. Another limitation is that the control group had colon radiographs but the cases pelvic radiographs. A study comparing urograms with pelvic radiographs found that joint space width was on average 10% greater on the urograms29, while another study found no significant influence of beam angle23. If such an effect exists it could introduce a bias away from the null. On the other hand our results were also significant when using the Kellgren and Lawrence grading system, which is less dependent on joint space.

Patients who undergo colon radiography are not a random sample of the population. Subjects with hip OA, who are seen within health care more often, may more likely be referred to colon radiography than the background population, introducing selection bias away from the null. We do not have information on BMI. Higher BMI is linked to lower hip fracture risk30, and higher hip OA risk28,31. Obesity is linked to colon cancer32 and these subjects may more commonly undergo colon radiography, but we are not aware of any studies on the BMI of the average patient undergoing colon radiography. If our controls overall had a higher BMI than the fracture cases, then they might have had a higher risk for hip OA, compared to the background population introducing bias away from the null.

Due to the relatively long period of time that the cases and controls were sampled there could be a birth cohort effect. This might be, for example, due to the fact that BMI has increased steadily between birth cohorts. The mean birth year for our cases was 1918.5 (SD 9.7) and for our controls the mean birth year was 1922.5 (SD 7.8), so a birth cohort effect should affect our cases and controls equally.

A prerequisite for a hip fracture is a fall, which can be influenced by age, comorbidity and medications. OA increases the risk for falls33,34, which would in contrast to the above introduce bias towards the null. The sum and direction of the aforementioned biases and any unforeseen biases is difficult to ascertain.

We used an AP pelvic radiograph of patients who were admitted to our hospital because of a hip fracture for measurement of MJS and assessment of hip OA by Kellgren and Lawrence grade. We are not aware of previous publications where MJS measurements of the fractured hip have been used. We considered the possibility that the traumatised hip would not be representative of its normal state, perhaps due to bleeding or muscle spasm. We therefore validated this method. In 46 out of 50 cases the determination of presence of OA according to MJS was the same. The fact that in 4 out of 50 cases we got a different reading is not greater than would be expected from any re-reading of a radiograph. Hip adduction-abduction has also been shown to result in a mean difference of less than 0.2 mm joint space23, which we suggest would not significantly influence our interpretations. To evaluate the whether the relationship between hip OA and hip fracture is systemic or local to the arthritic hip we graded both the fractured hip and the contralateral hip.

The Kappa values for the fractured hip were in general acceptable and similar to previously published studies. The Kappa values for the contralateral hip were somewhat lower. The films used to evaluate reliability of the readings were chosen to include the full range of radiographic features for the fractured side, but as we used the same films for the contralateral side, the spread between different Kellgren and Lawrence grades was not even and we therefore believe that this difference may be the result of the inherent flaw of the Kappa method in such cases35.

Definition of hip OA varies between studies. Recent studies have criticised the use of Kellgren and Lawrence classification27, while others have shown that measurement of joint space width is reliable and reflects clinical status36. We used both joint space width and Kellgren and Lawrence classification of hip OA to facilitate comparison with other studies.

Data on possible secondary osteoporosis was only available for our fractured patients, ideally we would have had data on secondary osteoporosis for both cases and controls. Thus, we were not able to fully evaluate to what extent secondary osteoporosis influences our conclusions, but our results indicate that secondary osteoporosis is overrepresented amongst patients with hip fracture and hip OA.

Some studies suggest that an inverse relationship between OA and hip fracture exists4-6,16,37, while and others refute it7,8. Most of these studies are of case-control design, while the two studies that refute this relationship are prospective cohort studies. One might therefore suggest that the evidence is stronger that there is no such relationship, even though the studies that support it are greater in number. In fact, one cannot even assume that these studies oppose one another, because of the differences in case definitions in these studies. In cohort studies the exposure is determined at the start of the study. In the case of OA, the baseline radiographic examination in a cohort study does not give an accurate estimate of the prevalence of radiographic hip OA at the time of fracture, which can be several years after the baseline examination. Reports on knee OA have also supported an inverse relationship between OA and osteoporosis38 and a molecular basis and common pathophysiology was proposed for the inverse relationship between OA and osteoporosis39. A genetic component to both osteoporosis40 and OA41, may explain why these conditions seldom coexist.

Conclusions
To the best of our knowledge this is the first study based on radiographically verified hip OA to quantify the risk of hip OA in patients with hip fracture, and also the first study to address the possible influence of secondary osteoporosis. We found that patients with hip fracture have one-third the risk for having hip OA in the fractured hip when compared to controls having had colon radiography. In the contralateral hip, the risk decrease was similar for men, and slightly less for women.

Competing interests
The authors declare that they have no competing interests.

Authors’ contributions
JF, TI and SL planned the study. JF read radiographs of hip fractures, TI read radiographs of controls. JF collected data and did the statistical analysis. JF drafted the manuscript. TI, SL and ME revised the manuscript. All authors took part in analysing the findings and all authors approved the final version of the manuscript.

Acknowledgements
Supported by: Scientific Foundation of Akureyri Central Hospital, The Swedish Research Council (medicine), Lund University Medical Faculty, Region Skåne, the King Gustaf V 80-year Fund, and The Swedish Rheumatism Association.

References
  1. Astrom J, Beertema J: Reduced risk of hip fracture in the mothers of patients with osteoarthritis of the hip. J Bone Joint Surg Br 1992, 74:270-271.
  2. Biyani A, Simison AJ, Klenerman L: Intertrochanteric fractures of the femur and osteoarthritis of the ipsilateral hip. Acta Orthop Belg 1995, 61:83-91.
  3. Dequeker J, Goris P, Uytterhoeven R: Osteoporosis and osteoarthritis (osteoarthrosis). Anthropometric distinctions. JAMA 1983, 249:1448-1451.
  4. Cumming RG, Klineberg RJ: Epidemiological study of the relation between arthritis of the hip and hip fractures. Annals of the rheumatic diseases 1993, 52:707-710.
  5. Dequeker J, Johnell O: Osteoarthritis protects against femoral neck fracture: the MEDOS study experience. Bone 1993, 14(Suppl 1):S51-56.
  6. Foss MV, Byers PD: Bone density, osteoarthrosis of the hip, and fracture of the upper end of the femur. Annals of the rheumatic diseases 1972, 31:259-264.
  7. Arden NK, Nevitt MC, Lane NE, Gore LR, Hochberg MC, Scott JC, Pressman AR, Cummings SR: Osteoarthritis and risk of falls, rates of bone loss, and osteoporotic fractures. Study of Osteoporotic Fractures Research Group. Arthritis Rheum 1999, 42:1378-1385.
  8. Jones G, Nguyen T, Sambrook PN, Lord SR, Kelly PJ, Eisman JA: Osteoarthritis, bone density, postural stability, and osteoporotic fractures: a population based study. J Rheumatol 1995, 22:921-925.
  9. Colhoun EN, Johnson SR, Fairclough JA: Bone scanning for hip fracture in patients with osteoarthritis: brief report. J Bone Joint Surg Br 1987, 69:848.
  10. Wand JS, Hill ID, Reeve J: Coxarthrosis and femoral neck fracture. Clinical orthopaedics and related research 1992, 88-94.
  11. Arden NK, Griffiths GO, Hart DJ, Doyle DV, Spector TD: The association between osteoarthritis and osteoporotic fracture: the Chingford Study. Br J Rheumatol 1996, 35:1299-1304.
  12. Wolf O, Strom H, Milbrink J, Larsson S, Mallmin H: Differences in hip bone mineral density may explain the hip fracture pattern in osteoarthritic hips. Acta Orthop 2009, 1-6.
  13. McClung MR, Geusens P, Miller PD, Zippel H, Bensen WG, Roux C, Adami S, Fogelman I, Diamond T, Eastell R, et al.: Effect of risedronate on the risk of hip fracture in elderly women. Hip Intervention Program Study Group. N Engl J Med 2001, 344:333-340.
  14. Antoniades L, MacGregor AJ, Matson M, Spector TD: A cotwin control study of the relationship between hip osteoarthritis and bone mineral density. Arthritis Rheum 2000, 43:1450-1455.
  15. Styrkarsdottir U, Halldorsson BV, Gretarsdottir S, Gudbjartsson DF, Walters GB, Ingvarsson T, Jonsdottir T, Saemundsdottir J, Center JR, Nguyen TV, et al.: Multiple genetic loci for bone mineral density and fractures. N Engl J Med 2008, 358:2355-2365.
  16. Weintroub S, Papo J, Ashkenazi M, Tardiman R, Weissman SL, Salama R: Osteoarthritis of the hip and fracture of the proximal end of the femur. Acta orthopaedica Scandinavica 1982, 53:261-264.
  17. Ingvarsson T, Hägglund G, Jonsson H Jr, Lohmander LS: Incidence of total hip replacement for primary osteoarthrosis in Iceland 1982-1996. Acta orthopaedica Scandinavica 1999, 70:229-233.
  18. Hilliquin P, Pessis E, Coste section sign J, Mauget D, Azria A, Chevrot A, Menkes CJ, Kahan A: Quantitative assessment of joint space width with an electronic caliper. Osteoarthritis Cartilage 2002, 10:542-546.
  19. Kellgren JH, Lawrence JS: Radiological assessment of osteo-arthrosis. Annals of the rheumatic diseases 1957, 16:494-502.
  20. Rüedi TP, Buckley RE, Moran CG: AO Principles of Fracture Management. 2nd edition. AO Publishing; 2007.
  21. Ingvarsson T, Hägglund G, Lindberg H, Lohmander LS: Assessment of primary hip osteoarthritis: comparison of radiographic methods using colon radiographs. Annals of the rheumatic diseases 2000, 59:650-653.
  22. Croft P, Cooper C, Wickham C, Coggon D: Defining osteoarthritis of the hip for epidemiologic studies. Am J Epidemiol 1990, 132:514-522.
  23. Goker B, Sancak A, Haznedaroglu S, Arac M, Block JA: The effects of minor hip flexion, abduction or adduction and x-ray beam angle on the radiographic joint space width of the hip. Osteoarthritis Cartilage 2005, 13:379-386.
  24. Kelman A, Lane NE: The management of secondary osteoporosis. Best practice & research 2005, 19:1021-1037.
  25. Fitzpatrick LA: Secondary causes of osteoporosis. Mayo Clinic proceedings 2002, 77:453-468.
  26. Curtin LR, Klein RJ: Direct standardisation (age-adjusted death rates). Healthy People 2000 Stat Notes 1995, 1-10.
  27. Jacobsen S, Sonne-Holm S: Increased body mass index is a predisposition for treatment by total hip replacement. Int Orthop 2005, 29:229-234.
  28. Franklin J, Ingvarsson T, Englund M, Lohmander LS: Sex differences in the association between body mass index and total hip or knee joint replacement resulting from osteoarthritis. Annals of the rheumatic diseases 2009, 68:536-540.
  29. Auleley GR, Duche A, Drape JL, Dougados M, Ravaud P: Measurement of joint space width in hip osteoarthritis: influence of joint positioning and radiographic procedure. Rheumatology (Oxford) 2001, 40:414-419.
  30. De Laet C, Kanis JA, Oden A, Johanson H, Johnell O, Delmas P, Eisman JA, Kroger H, Fujiwara S, Garnero P, et al.: Body mass index as a predictor of fracture risk: a meta-analysis. Osteoporos Int 2005, 16:1330-1338.
  31. Lohmander LS, Gerhardsson de Verdier M, Rollof J, Nilsson PM, Engstrom G: Incidence of severe knee and hip osteoarthritis in relation to different measures of body mass: a population-based prospective cohort study. Annals of the rheumatic diseases 2009, 68:490-496.
  32. Caan BJ, Coates AO, Slattery ML, Potter JD, Quesenberry CP Jr, Edwards SM: Body size and the risk of colon cancer in a large case-control study. Int J Obes Relat Metab Disord 1998, 22:178-184.
  33. Nevitt MC, Cummings SR, Kidd S, Black D: Risk factors for recurrent nonsyncopal falls. A prospective study. JAMA 1989, 261:2663-2668.
  34. Campbell AJ, Borrie MJ, Spears GF: Risk factors for falls in a community-based prospective study of people 70 years and older. J Gerontol 1989, 44:M112-117.
  35. Byrt T, Bishop J, Carlin JB: Bias, prevalence and kappa. J Clin Epidemiol 1993, 46:423-429.
  36. Dougados M, Gueguen A, Nguyen M, Berdah L, Lequesne M, Mazieres B, Vignon E: Radiological progression of hip osteoarthritis: definition, risk factors and correlations with clinical status. Annals of the rheumatic diseases 1996, 55:356-362.
  37. Vestergaard P, Rejnmark L, Mosekilde L: Osteoarthritis and risk of fractures. Calcif Tissue Int 2009, 84:249-256.
  38. Hart DJ, Mootoosamy I, Doyle DV, Spector TD: The relationship between osteoarthritis and osteoporosis in the general population: the Chingford Study. Annals of the rheumatic diseases 1994, 53:158-162.
  39. Lories RJ, Peeters J, Bakker A, Tylzanowski P, Derese I, Schrooten J, Thomas JT, Luyten FP: Articular cartilage and biomechanical properties of the long bones in Frzb-knockout mice. Arthritis Rheum 2007, 56:4095-4103.
  40. Spector TD, Keen RW, Arden NK, Morrison NA, Major PJ, Nguyen TV, Kelly PJ, Baker JR, Sambrook PN, Lanchbury JS, et al.: Influence of vitamin D receptor genotype on bone mineral density in postmenopausal women: a twin study in Britain. BMJ 1995, 310:1357-1360.
  41. Ingvarsson T, Stefansson SE, Hallgrimsdottir IB, Frigge ML, Jonsson H Jr, Gulcher J, Jonsson H, Ragnarsson JI, Lohmander LS, Stefansson K: The inheritance of hip osteoarthritis in Iceland. Arthritis Rheum 2000, 43:2785-2792.


top

top of page     readers comments     contact us     webmaster     legal notices     home     subscribe
Current Issue Orthopaedic Products SurgeonSpeak Future Events Articles Company Info OPNews
Copyright 2012 © Pelican Magazines Ltd  -  developed by netdirect sales

a Pelican Group Company